Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

ESP-16HD/32HD/64HD Specifications

  return to ESP Scanners main product page

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 

 

 
Parameter
DTC

CONVENTIONAL

Units
Comments
32HD 64HD 16HD 32HD 64HD
PNEUMATICS
Number of Pressure Inputs 32 64 16 32 64 ports  
Pressure Span (1) ±10 (2.5), ±20 (5.0)
±1(7), ±2.5(17), ±5(34), ±10(69), ±15(103),
±30(207), ±45(310), ±50(345), +100(689)
inWC (kPa)
psid (kPa)
psid (kPa)
 
Proof Pressure (2) 400
150
%FS
%FS
< or = to 30 psid (207 kPa)
> 30 psid (207 kPa)
Max Reference Pressure
(Line Pressure)
50 (345)
15 (103)
psid (kPa)
psid (kPa)
> or = to 2.5 psid (17 kPa)
less than 2.5 psid (17 kPa)
Manifold Control Pressure 100 (689)
125 (862)
psig (kPa)
psig (kPa)
min                           requires only
max                   momentary pulse
Optional Purge Pressure greater than max input pressure
150 (1050)
psig (kPa)
psig (kPa)
min
max
Media:
  Measurement Ports

  Reference Ports

non-corrosive gases up to 100%
non-condensing humidity
dry, non-corrosive gases
   
Pneumatic Connection .040 or .063 .040 .040 or .063 .040 or .063 .040

inches OD

SS bulged tubulation
PERFORMANCE
Static Accuracy (3, 4)
  @ 100% range


  @ 33% range (7)

  2nd order correction (5)


  4th order correction (6)
 

±0.03
±0.06

±0.05

N/A
N/A

N/A
N/A


N/A
N/A

N/A

±0.15
±0.20

±0.05
±0.10

% FS
% FS

% FS

% FS
% FS

% FS
% FS

after span cal(ranges > or = to 5 psid)
after span cal(ranges < 5 psid)

after span cal(ranges > or = to 5 psid)

> 2.5 psid (17 kPa)
< or = to 2.5 psid (17 kPa)

> 2.5 psid (17 kPa)
< or = to 2.5 psid (17 kPa)
Thermal Zero Error N/A ±0.06
±0.20

%FS/șC
%FS/șC

> 2.5 psid (17 kPa)
< or = to 2.5 psid (17 kPa)
Thermal Span Error N/A ±0.02
±0.03

%FS/șC
%FS/șC

> 2.5 psid (17 kPa)
< or = to 2.5 psid (17 kPa)
Total Thermal Stability
  @ 100% range

  @ 33% range (7)

±0.002
±0.004

±0.003

N/A


N/A


%FS/șC
%FS/șC

%FS/șC


> or = to 5 psid (34 kPa)
less than 5 psid (34 kPa)

> or = to 5 psid (34 kPa)
Line Pressure Effect

±0.001
±0.002

%FS/psi
%FS/psi

> 2.5 psid (17 kPa)
< or = to 2.5 psid (17 kPa)
ELECTRICAL
Power Supply
  +5 VDC
  +12 VDC
  -12 VDC

30
60
20

30
85
20

25
5
3

55
5
3

75
5
3


mA
mA
mA


max
max
max
Full Scale Output

±3.5 - 4.5

VDC

optional outputs available
Port Selection 5 6 4 5 6

bit

0-12 VDC, CMOS Logic
Scan Rate

20,000

chs/sec

Electrical Connection

15 pin subminiature D-shell (8)

  ITT Cannon (PN:MDM15SSB)
ENVIRONMENTAL/PHYSICAL
Comp Temp Range 0 to 80
șC
Operating Temp Range -25 to 80 șC  
Storage Temp -25 to 100 șC  
Weight 3.9
(112)
5.1
(143)
1.6
(46)
3.7
(106)
4.9
(140)
oz.
(g)
 
HEATER CHAMBERS
Setpoint Adjustability ±3
șC
 
Thermal Compensation N/A (9) 10:1   reduction in the effects of temperature to the scanner
Ambient Temperature

-55

șC
min
Power Requirements

28 VDC @ 1.5A

  max
Electrical Connection

Two leads unterminated

  Red - +28 VDC
Black - GND
Temperature Overload 90 șC with Auto Reset
Weight 6.8
(192)
6.8
(192)
5.6
(157)
6.8
(192)
6.8
(192)

oz.
(g)

 
(1) Consult factory for pressure ranges not listed.
(2) Maximum applied pressure without reducing sensor performance.
(3) Static accuracy includes combined errors due to non-linearity, hysteresis, and non-repeatability.
(4) Scanner span calibration recommended at ambient temperature every 6 months to maintain static accuracy specification.
(5) 2nd order equation of the form Px = C0 + C1 (Vx) + C2 (Vx)2
(6) 4th order equation of the form Px = C0 + C1 (Vx) + C2 (Vx)2 + C3 (Vx)3 + C4 (Vx)4
(7) Accuracy is expressed as a percent of the deranged Full Scale.
(8) ESP-16HD supplied with intermediate cable terminating with 15-pin subminiature D-shell.
(9) Heater chambers for DTC Series scanners are only required to maintain minimum operating temperature.


Specifications subject to change without notice.


 

 

 
 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search