Pressure Systems rackmount version of its highly successful NetScanner System features the latest generation Model 98RK-1 Scanner Interface Rack and Model 9816 Pneumatic Intelligent Pressure Scanners for ultra high speed scanning of multiple pneumatic measurements in turbomachinery testing.
|

Series 9816/98RK-1 Features
- ±0.05% Accuracy
- 100 Hz EU Throughput
- Autonegotiating 10/100/1G Base T Ethernet
- TCP and UDP Protocol
- Pressure Ranges up to 850 psid (5860 kPa)
- Rezero, Purge and Leak Check Functionality
- Triple Scan List Capability
- Hardware and Software Initiated Data Acquisition
Series 9816/98RK Applications
- Turbine Engine Research
- Turbomachinery Test Stands
|
|
|
The enhanced Model 98RK-1 rack utilizes a high performance switch enabling the 9816 scanners to communicate with the host computer via an auto-negotiating 10/100/1G Base T Ethernet interface. The rack also provides power and pneumatic connections for up to eight internally-housed 9816 scanners. Additionally, the 98RK-1 can provide power and communication for up to three external NetScanner System components such as PSI’s Model 9022 for all-media pressure measurements or Model 9032 for high-accuracy barometric or reference pressure measurements.
Model 9816 scanners integrate 16 silicon piezoresistive pressure sensors of any pressure range with a microprocessor. Each field-replaceable pressure sensor incorporates a temperature sensor and an integral EEPROM for storage of personalized calibration data as well as sensor information such as serial number, pressure range, factory calibration date, and user-managed last or next calibration date. Integrating the EEPROM within the sensor enables simplified field replacement of sensors with automatic uploading of sensor data during power up.
The microprocessor uses the sensor data to correct for zero, span, linearity and thermal errors. It also performs digital temperature compensation to reduce sensor thermal errors by a factor of ten or more over conventional compensation. The microprocessor also controls the actuation of an integral manifold to execute on-line zero calibrations upon request. This capability virtually eliminates sensor zero drift error and provides guaranteed system accuracy of up to ±0.05% FS (Full Scale) after rezero.
Each 9816 can sample using up to three concurrent scan lists at continuous rates up to 100 measurements per channel per second in engineering units. Pressure data is output over the Ethernet interface supporting both TCP and UDP protocols. The NetScanner System is supplied with comprehensive Visual Basic-based software called NUSS for PC compatible computers. Firmware upgrades are made available at no charge via PSI’s website and can be downloaded to the 9816 scanners using NUSS.
The NetScanner™ System is an array of system components measuring a variety of parameters and sharing the same command set that can be networked to form a comprehensive distributed intelligent data acquisition system.
Dated 06/08
| Parameter |
9816 |
98RK-1 |
Units |
Comments |
|
PNEUMATICS
|
| Pressure Ranges (1) |
10 (2.5)
20 (5.0)
|
N/A |
inches WC
(kPa) |
ranges < 100 psi (689 kPa) are bidirectional
ranges >= 100 psi (689 kPa) are unidirectional
|
|
1 (7.0)
2.5 (17)
5 (34)
10 (69)
15 (103)
30 (207)
45 (310)
100 (689)
250 (1724)
500 (3447)
750 (5171)
850 (5860)
|
N/A |
psid (kPa) |
Number of
Measurement Inputs |
16 |
N/A |
|
|
| Pneumatic Fittings |
.063 bulged tubulation, or
SAE 5/16-24 o-ring boss
(if front access specified) |
SAE 7/16-24
o-ring boss |
|
variety of compression
adapter fittings available
See page 6 for standard fittings |
| Media |
Dry, non-corrosive gas |
|
for liquid pressures refer to
Model 9022 datasheet |
| Proof Pressure (2) |
3.0x
2.0X
1.5X |
N/A |
FS min |
range < 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range > 100 psi (689 kPa) |
| Burst Pressure |
5.0X
3.0X
2.0X |
N/A |
FS min |
range < 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range > 100 psi (689 kPa) |
| Max Line Pressure (3) |
100 (689)
600 (4134) |
N/A |
psi (kPa)
psi (kPa) |
range <= 5 psi (34 kPa)
range > 5 psi (34 kPa) |
Manifold Control
Pressure |
80 (552)
125 (862)
|
psi (kPa) |
min
max |
| Purge Pressure (4) |
greater max input pressure
850 (5860)
|
psi (kPa)
psi (kPa) |
min
max |
|
STATIC PERFORMANCE
|
| Measurement Resolution |
±0.003 |
N/A |
% FS |
|
Static Accuracy (5)
After Rezero |
±0.05
±0.15 |
N/A |
% FS
% FS |
range > 2.5 psi (17 kPa)
range <= 2.5 psi (17 kPa) |
| Total Thermal Error (7) |
±0.001
±0.0015 |
N/A |
%FS/ºC
%FS/ºC |
range > 2.5 psi (17 kPa)
range <= 2.5 psi (17 kPa) |
| Line Pressure Effect |
±0.0003
±0.0008
±0.001 |
N/A |
% FS/psi
% FS/psi
% FS/psi |
range <= 100 psi (689 kPa)
range >5 psi (45 kPa) <100 psi (689 kPa)
range <= 5 psi (34 kPa) |
| Measurement Rate |
100 |
meas/ch/sec |
engineering units |
|
COMMUNICATION
|
| Ethernet |
10BaseT
half duplex |
10/100/1000 BaseT
auto negotiating
|
|
|
| Protocol |
TCP and UDP |
% FS
|
static or dynamic IP addressing |
|
ELECTRICAL
|
| Input Voltage Options |
N/A |
85-264 |
VAC |
47-63 Hz |
| Operating Power |
N/A |
65 (7) |
watts |
|
| Hardware Trigger |
2.5
|
VDC |
TTL compatible differential input, ±5 VDC common mode voltage |
|
ENVIRONMENTAL/PHYSICAL
|
| Calibrated Temp Range (10) |
0 to 60 |
ºC |
|
| Operating Temp Range |
0 to 50 |
ºC |
up to 95% non-condensing humidity |
| Storage Temp |
-20 to 80 |
ºC |
|
| Size |
2.06 x 7.00 x 15.25
(5.2 x 17.7 x 38.7) |
17.00 x 8.75 x 19.00
(43.2 x 22.2 x 48.3) |
inches
(cm) |
L x W x H |
| Weight |
6.5
(3.0) |
26.9
(12.2) |
lb
(kg) |
|
- Consult Factory for intermediate pressure ranges and multirange configurations.
- Maximum applied pressure without reducing sensor performance.
- For anticipated Line (Reference) Pressures . . . above 100 psi (689 kPa) on FS Pressure Ranges of ,= 5 psi (34 kPa) or
- above 600 psi (4137 kPa) on FS Pressure Ranges of > 5 psi (34 kPa)
- calibration manifold should not be shifted at line pressure
- rate of line pressure change should be no more than 10 psi/sec
- Purge pressure must be greater than maximum input pressure.
- Static accuracy includes combined errors due to non-linearity, hysteresis and non-repeatability.
- Includes thermal effects of zero & span relative to 25ºC.
- Maximum power consumption with full compliment of 9816 scanners and external NetScanner System components in highest power draw configurations.
Specifications subject to change without notice.
Updates Errata for 98RK User's Manual, August 2001
Page 1-6
Section 1.3.2 Manifolds and Pressure Connections
Add the following NOTE to the paragraph:
"9816 scanners with front-mount quick disconnect (QDC) panels have all necessary input fittings on the panel. Therefore, the run reference (RUNREF) input ports on the rear of the 98RK are not utilized. Previously, compression fittings were installed in these rear ports. The input ports now have plugs, rather than fittings, installed.
"Page 5-1 Figure 5.1 (Exploded View of 9816 Calibration Manifold)
Replace the drawing on Page 5-1 (Figure 5.1) with the following:

Figure 5.1
Exploded View of 9816 Calibration Manifold
Page 5.10
Paragraph (4) should read;
Slide the front panel forward as far as it will go (approximately d").
Page 5-16Section 5.1.6.2 Tubing Plate O-Ring ReplacementAdd a new paragraph (5) to read as follows:
"Inspect for the presence of optional shim washers around each of the six (6) screws. If washers are present, retain for use in reassembly."
Renumber remaining paragraphs accordingly.Change paragraph (7) to read: (additions are in italics)
Examine the tubing plate and calibration manifold to verify that no contaminants are on either surface. This generally requires microscopic examination. Replace the tubing plate by slowly placing the plate on the calibration manifold housing. Make sure the O-ring/cup seal side is down toward the pneumatic sliding manifold and the guiding pin on the calibration manifold housing fits into the mating hole of the tubing plate. Also, ensure that shim washers, if used, are reinstalled between the tubing plate and the calibration manifold assembly.
Page 5-20
Section 5.1.6.6 Module Rear Manifold O-Ring Replacement
Paragraph (4a) to read as follows:For glued O-ring replacement (O-rings will be black)
(4a) There are only two O-rings on this panel that are glued; the larger black ones. One is for the purge port and the other surrounds a blanking pin. Using a glue syringe, line the O-ring cup(s) with appropriate glue, such as 3M Scotch Grip 1357. Insert a new O-ring (PSI part number 61-03-50120) and press it into the cup with your hand. Allow the glue to dry for approximately 10-15 minutes. Clean any excess glue with appropriate cleaning solvent, and dry. Be extremely careful not to allow any excess glue to cover the tubing ports.
Pages 5-21 and 5-22
Section 5.1.6.7 Rear Panel Quick Disconnect (QDC) O-Ring Replacement
NONE of the O-rings on this panel are glued. All are pressed into place over a flared center tube.
PSI part number for these O-rings is 61-03-4762. All O-rings are black.
Delete paragraph (6) and renumber the remaining paragraphs (6) through (10).
Pages 5-22 and 5-23
Section 5.1.6.8 Front Panel Tubing Plate Quick Disconnect (QDC) O-Ring Replacement
NOTE: NONE of the O-rings on this plate are glued. All O-rings are pressed into place over a flared
center tube. PSI part number for these O-rings is 61-03-4762. All O-rings are black. |
Delete paragraph (4a)"For glued O-ring replacement..."
Renumber paragraph (4b) to (4a), and delete the words "For all" following the new paragraph (4a
Netscanner™ System: 9816 Version 2.42 Release Notes
Last Released Version: Version 2.35
Version 2.42 Changes
- Changed default factory subnet mask from 128.0.0.0 to 192.0.0.
- Increased buffers available for use by the network stack.
- Address I2C issues with microcontroller mask L75M (affects new products only).
- General cleanup.
Command Summary: None.
Command Change Summary: None