You are here: Products  » Industrial » NetScanner™ System » Series 9400

  Search
NetScanner™ Series 9400 Digitally Compensated All-Media Pressure Transducer   

Series 9400 are highly accurate, digitally compensated pressure transducers designed specifically for use with PSI’s Model 9021 or 9022 All-Media Intelligent Pressure Scanners in applications requiring compatibility with liquid and high-pressure media. The 9400 utilizes a highly accurate pressure sensor assembly specifically designed for hostile fluids and gases.

Series 9400 Features

  • Embedded EEPROM-based Calibration Data
  • Digital Sensor Compensation
    -  Up to ±0.05% FS Static Accuracy
    -  Up to ±0.005%/ºC Thermal Stability
  • Pressure Ranges: 0-5 psi (34 kPa)
       to 0-10000 psi (69 MPa)
  • Welded 316 SS Construction

Series 9400 Applications

  • Turbomachinery Test Stands
  • Hydraulic/Pneumatic Systems
  • Process Control
  • Environmental Monitoring
Support

Updates

Application Notes

Software/Firmware

FAQs

PDF Downloads

Datasheet

Certifications

 

The assembly is integrated with supporting electronics in a durable waterproof housing constructed of 316 SS. Standard pressure ranges are available from 0-5 to 0-10,000 psi in gage, absolute and differential reference pressure formats.

Series 9400 transducers achieve high accuracy and thermal stability through the use of digital compensation technology to correct zero, span, and linearity errors over the operating pressure and temperature range. Each digitally compensated transducer contains an integral serial EEPROM storing factory generated calibration data. This data is uploaded into the Model 9021 or 9022 Intelligent Pressure Scanners upon power-up and used to compensate for the inherent transducer errors during use. Static accuracy of up to ±0.05% FS with thermal stability up to ±0.005%/ºC are achieved using this technique. The Model 9021 and 9022 scanners support output pressure measurements from the 9400 transducers in engineering units over an Ethernet interface using TCP or UDP protocols.

The Series 9400 All-Media Pressure Transducer is one component of the NetScanner™ System. Multiple NetScanner components measuring a variety of parameters and sharing the same command set can be networked to form a distributed intelligent data acquisition system.

Dated 0209

Parameter

9400

9401

9402

Units

Comments

PNEUMATICS

Pressure Reference

Gauge (1)

Absolute

Differential

 

 
Pressure Ranges
9400
Gauge

5 (34)
10 (69)
15 (103)
50 (345)
100 (689)
250 (1724)
500 (3447)
750 (5171)
1500 (10342)
3000 (20684)
5000 (34474)
10000(68947)
9401
Absolute

15 (103)
30 (207)
50 (345)
100 (689)
250 (1724)
500 (3447)
750 (5171)
1500 (10342)
3000 (20684)
5000 (34474)
10000(68947)
9402
Differential

5 (34)
10 (69)
15 (103)
50 (345)
100 (689)
250 (1724)
psi (kPa)
Contact factory for other ranges.
 
Proof Pressure (2)
3.0
1.5
x FS
range less than or equal to 100 psi (689 kPa)
range greater than 100 psi (689 kPa)
Burst Pressure
5x
3x
2x
x FS
range equal to 5 - 500 psi
range greater than 500 - 3000 psi
range greater than 3000 psi

STATIC PERFORMANCE

Static Accuracy (3)
 
±0.05
±0.25

±0.50
±0.5
N/A

N/A
% FS
% FS

% FS
range less than 750 psi
range greater than 750 psi to less than equal to 3000 psi
range greater than 3000 psi
Total Thermal Error (4)
 

±0.005
±0.05

%FS/ºC
%FS/ºC

digital compensation (6)
analog compensation (7)
Thermal Hysteresis
±0.2
% FS after cycling over full temp range
Max Line Pressure N/A 1000 (6885) psi (kPa)  
Line Pressure Effect (5)
N/A
±0.01
%FS/psi  

ENVIRONMENTAL

Wetted Materials

316 SS & Viton

   
Compensated Temp Range

0 to 50
-30 to 40

ºC

standard
optional
Operating Temp Range

-30 to 100

ºC

 

ELECTRICAL

Excitation

12 - 36

VDC

 
Power Supply Rejection

±.001

%/VDC

 
Output
Vo
Vt

0 - 4.9
3.0 - 4.0

VDC
VDC

 
Output Impedance
Vo
Vt

1000
1000

Ohm
Ohm


max
max
Insulation Resistance

50

M Ohm

@ 50 VDC

PHYSICAL/ENVIRONMENTAL

Acceleration
±0.02

±0.01
% FS/g

% FS/g
Range less than equal to 15 psi (103 kPa)
Range greater or equal to 30 psi (207 kPa)
Vibration
±0.05
% FS/g 30 g peak
10 Hz - 2kHz
Weight
6 (170)
17 (480)
oz. (gm)
 
Pressure Connection
1/4" NPT, Male
AN4, Male
G 1/4, Male
M14, Male
7/16"
Female SAE
o-ring Boss
   
Electrical Connection
Standard
Optional
PTIH-12-8P Bendix
PVC Jacketed Cable
   
  1. Pressure ranges greater than 750 psi are "sealed" gage rather than "vented" gage.
  2. Maximum pressure which can be applied without causing calibration shift.
  3. Static accuracy includes the combined errors due to nonlinearity, hysteresis and nonrepeatability following a zero calibration of the sensor.
  4. Includes effects of zero and span relative to 25ºC.
  5. Primarily zero offset.
  6. For digital compensation type transducers, the on-board EEPROM contains full operating range temperature and nonlinearity correction coefficients.
  7. For analog compensation type transducers, the on-board EEPROM contains room temperature (approx. 25ºC) nonlinearity correction coefficients.

Specifications subject to change without notice