The Model 9116 Pneumatic Intelligent Pressure Scanner is a completely self-contained high performance pressure acquisition module for multiple measurements of dry, non-corrosive gases. The scanner integrates 16 silicon piezoresistive pressure sensors of any pressure range with the latest microprocessor technology in a compact, rugged package.

Model 9116 Features

  • User Accessible Memory for Test Configuration Management
  • ±0.05% FS System Accuracy
  • EU Throughput Rates of 500 Hz
  • Auto-Negotiating 10/100 BaseT Ethernet with TCP & UDP Protocol
  • Pressure Ranges up to 850 psid (5860 kPa)
  • Built-in Rezero, Purge and Leak Check Features
  • Triple Scan List Capability
  • Rugged Splashproof Enclosure
  • Hardware and Software Triggered Data Acquisition

Model 9116 Applications

  • Turbomachinery Test Stands
  • Windtunnel Wall Pressure Measurements

 

Support

Updates

Application Notes

Software/Firmware

FAQs

PDF Downloads

Datasheet

Manual

Certifications

 

Each field-replaceable pressure sensor incorporates a temperature sensor and EEPROM for storage of calibration data as well as sensor identification information such as pressure range, factory calibration date, and user-managed last or next calibration date.

The microprocessor uses the data from the EEPROM to correct for sensor zero, span, linearity and thermal errors. It also performs digital temperature compensation of the sensors to reduce thermal errors by a factor of ten or more over conventional sensor compensation. The microprocessor also controls the actuation of an integral manifold to execute on-line zero or span calibration upon request. This capability virtually eliminates sensor zero drift error and provides guaranteed system accuracy of up to ±0.05% FS (Full Scale) after rezero.

This enhanced pressure scanner provides User Accessible Memory (UAM) to facilitate Test Configuration Management (TCM) through functionality derived from the IEEE-1451 standard for smart transducers. This enables test parameters to be entered and stored within the scanner for access by the host controller during test setup and operation.

The 9116 can sample using up to three concurrent scan lists at continuous rates up to 500 measurements per channel per second in engineering units. Pressure data in engineering units is output through an auto-negotiating 10 or 100 Mbit Ethernet interface supporting both TCP and UDP protocols. The scanner is supplied with comprehensive Visual Basic-based software called NUSS for PC compatible computers. Firmware upgrades are made available at no charge via PSI’s website and can be downloaded to the scanner over the Ethernet interface using NUSS.

The Model 9116 Intelligent Pressure Scanner is one component of the NetScanner™ System. Multiple NetScanner components measuring a wide variety of parameters and sharing the same command set can be networked to form a distributed intelligent data acquisition system.

Dated 0904

Parameter

9116

Units
Comments

PNEUMATICS

Pressure Range (1)
10 (2.5)
20 (5.0)
In WC (kPa)
ranges greater than or equal to 100 psi (689 kPa) are unidirectional only.
1 (7)
2.5 ( 17)
5 (34)
10 (69)
15 (103)
30 (207)
45 (310)
100 (689)
250 (1724)
500 (3447)
750 (5171)
850 (5860)
psid (kPa)
Number of Pressure
Inputs
16

 

available in any combination of ranges
Pneumatic Fittings
1/16", 1/8", 1/4"
5/16" - 24 SAE
.063", .040" bulge tube
compression fittings
o-ring boss
QDC tubing plate
Media
Dry, non-corrosive gas

 

 
Proof Pressure (2)
3.0X
2.0X
1.5X
FS min
range less than 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range greater than 100 psi (689 kPa)
Burst Pressure
5.0X
3.0X
2.0X
FS min
range less than 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range greater than 100 psi (689 kPa)
Max Line Pressure (3)
 
100 (689)
600 (4137)
psi (kPa)
range less than or equal to 5 psi (34 kPa)
range greater than 5 psi (34 kPa)
Manifold Control
Pressure
80 (552)
125 (862)
psi (kPa)
min
max
Purge Pressure (4)
 
greater than max input pressure
850 (5860)
psi (kPa)
min
max
 

STATIC PERFORMANCE

Measurement
Resolution
±0.003
% FS
 
Static Accuracy (5)
After Rezero
±0.05
±0.15
% FS
% FS
range greater than 2.5 psi (17 kPa)
range less than or equal to 2.5 psi (17 kPa)
Total Thermal Error (6)
±0.001
±0.0015
%FS/ºC
%FS/ºC
range greater than 2.5 psi (17 kPa)
range less than or equal to 2.5 psi (17 kPa)
Line Pressure Effect
±0.0003
±0.0008

±0.001
% FS /psi
% FS/psi

% FS/psi
range greater or equal to 100 psi (689 kPa)
range greater than 5 psi (34 kPa) to
  less than 100 psi (689 kPa)
range less than or equal to 5 psi (34 kPa)
Measurement Rate
500
meas/ch/sec
 

COMMUNICATION

Ethernet
10/100 BaseT

 

auto negotiating, half duplex/full duplex
Protocol
TCP and UDP

 

static or dynamic IP addressing

ELECTRICAL

Input Voltage
18 - 36
VDC unregulated
Input Current
225
mA maximum @ 24 VDC
Hardware Trigger
Threshold
2.5
VDC TTL compatible differential
input, ±5 VDC common
mode voltage

ENVIRONMENTAL/PHYSICAL

Calibrated Temp
Range (7)
  Standard
  Optional
 

0 to 60
-30 to 60

 

ºC
ºC

 
Operating
Temp Range

  Standard
  Optional
 

-20 to 70
-30 to 80

 

ºC
ºC

 

up to 95% non-condensing humidity

Storage Temp
-30 to 80
ºC
 
Size
9.50 x 3.50 x 3.54
(24.1 x 8.9 x 8.98)
inches
(cm)
L x W x H
Weight
6.5
(2.95)
lb
(kg)
 
Splashproof Rating

NEMA 13

   
Acoustic Noise Rating

160dB

   
Shock Rating

30

g

 
  1. Standard ranges listed, consult factory for special ranges.
  2. Maximum applied pressure without reducing sensor performance.
  3. For anticipated Line (Reference) Pressures . . . above 100 psi (689 kPa) on FS Pressure Ranges of less than or equal to 5 psi (34 kPa) or
    • above 600 psi (4137 kPa) on FS Pressure Ranges of greater than 5 psi (34 kPa)
    • calibration manifold should not be shifted at line pressure
    • rate of line pressure change should be no more than 10 psi/sec
  4. Purge pressure must be greater than maximum input pressure.
  5. Static accuracy includes the combined errors due to non-linearity, hysteresis and non-repeatability.
  6. Includes thermal effects of zero & span relative to 25ºC.
  7. Consult factory for special temperature range calibrations.

Specifications subject to change without notice.

9116 Version 5.10 Release

Last Released Version: Version 5.09

Version 5.10 Changes

  1. Updated drivers to accommodate alternate memory parts.

   © 2010 Pressure Systems Inc. Terms Of Use | Privacy Statement