The Model 9116 Pneumatic Intelligent Pressure Scanner is a completely self-contained high performance pressure acquisition module for multiple measurements of dry, non-corrosive gases. The scanner integrates 16 silicon piezoresistive pressure sensors of any pressure range with the latest microprocessor technology in a compact, rugged package.
|

Model 9116 Features
- User Accessible Memory for Test Configuration Management
- ±0.05% FS System Accuracy
- EU Throughput Rates of 500 Hz
- Auto-Negotiating 10/100 BaseT Ethernet with TCP & UDP Protocol
- Pressure Ranges up to 850 psid (5860 kPa)
- Built-in Rezero, Purge and Leak Check Features
- Triple Scan List Capability
- Rugged Splashproof Enclosure
- Hardware and Software Triggered Data Acquisition
Model 9116 Applications
- Turbomachinery Test Stands
- Windtunnel Wall Pressure Measurements
|
|
|
Each field-replaceable pressure sensor incorporates a temperature sensor and EEPROM for storage of calibration data as well as sensor identification information such as pressure range, factory calibration date, and user-managed last or next calibration date.
The microprocessor uses the data from the EEPROM to correct for sensor zero, span, linearity and thermal errors. It also performs digital temperature compensation of the sensors to reduce thermal errors by a factor of ten or more over conventional sensor compensation. The microprocessor also controls the actuation of an integral manifold to execute on-line zero or span calibration upon request. This capability virtually eliminates sensor zero drift error and provides guaranteed system accuracy of up to ±0.05% FS (Full Scale) after rezero.
This enhanced pressure scanner provides User Accessible Memory (UAM) to facilitate Test Configuration Management (TCM) through functionality derived from the IEEE-1451 standard for smart transducers. This enables test parameters to be entered and stored within the scanner for access by the host controller during test setup and operation.
The 9116 can sample using up to three concurrent scan lists at continuous rates up to 500 measurements per channel per second in engineering units. Pressure data in engineering units is output through an auto-negotiating 10 or 100 Mbit Ethernet interface supporting both TCP and UDP protocols. The scanner is supplied with comprehensive Visual Basic-based software called NUSS for PC compatible computers. Firmware upgrades are made available at no charge via PSI’s website and can be downloaded to the scanner over the Ethernet interface using NUSS.
The Model 9116 Intelligent Pressure Scanner is one component of the NetScanner™ System. Multiple NetScanner components measuring a wide variety of parameters and sharing the same command set can be networked to form a distributed intelligent data acquisition system.
Dated 0904
| Parameter |
9116
|
Units |
Comments |
|
PNEUMATICS
|
| Pressure Range (1) |
10 (2.5)
20 (5.0) |
In WC (kPa) |
ranges greater than or equal to 100 psi (689 kPa) are unidirectional only. |
1 (7)
2.5 ( 17)
5 (34)
10 (69)
15 (103)
30 (207)
45 (310)
100 (689)
250 (1724)
500 (3447)
750 (5171)
850 (5860) |
psid (kPa) |
Number of Pressure
Inputs |
16 |
|
available in any combination of ranges |
| Pneumatic Fittings |
1/16", 1/8", 1/4"
5/16" - 24 SAE
.063", .040" bulge tube |
compression fittings
o-ring boss
QDC tubing plate |
| Media |
Dry, non-corrosive gas |
|
|
| Proof Pressure (2) |
3.0X
2.0X
1.5X |
FS min |
range less than 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range greater than 100 psi (689 kPa) |
| Burst Pressure |
5.0X
3.0X
2.0X |
FS min |
range less than 15 psi (103 kPa)
range 15-100 psi (103-689 kPa)
range greater than 100 psi (689 kPa) |
Max Line Pressure (3)
|
100 (689)
600 (4137) |
psi (kPa) |
range less than or equal to 5 psi (34 kPa)
range greater than 5 psi (34 kPa) |
Manifold Control
Pressure |
80 (552)
125 (862) |
psi (kPa) |
min
max |
Purge Pressure (4)
|
greater than max input pressure
850 (5860) |
psi (kPa) |
min
max
|
|
STATIC PERFORMANCE
|
Measurement
Resolution |
±0.003 |
% FS |
|
Static Accuracy (5)
After Rezero |
±0.05
±0.15 |
% FS
% FS |
range greater than 2.5 psi (17 kPa)
range less than or equal to 2.5 psi (17 kPa) |
| Total Thermal Error (6) |
±0.001
±0.0015 |
%FS/ºC
%FS/ºC |
range greater than 2.5 psi (17 kPa)
range less than or equal to 2.5 psi (17 kPa) |
| Line Pressure Effect |
±0.0003
±0.0008
±0.001 |
% FS /psi
% FS/psi
% FS/psi |
range greater or equal to 100 psi (689 kPa)
range greater than 5 psi (34 kPa) to
less than 100 psi (689 kPa)
range less than or equal to 5 psi (34 kPa) |
| Measurement Rate |
500 |
meas/ch/sec |
|
|
COMMUNICATION
|
| Ethernet |
10/100 BaseT |
|
auto negotiating, half duplex/full duplex |
| Protocol |
TCP and UDP |
|
static or dynamic IP addressing |
|
ELECTRICAL
|
| Input Voltage |
18 - 36 |
VDC |
unregulated |
| Input Current |
225 |
mA |
maximum @ 24 VDC |
Hardware Trigger
Threshold |
2.5 |
VDC |
TTL compatible differential
input, ±5 VDC common
mode voltage |
|
ENVIRONMENTAL/PHYSICAL
|
Calibrated Temp
Range (7)
Standard
Optional |
0 to 60
-30 to 60
|
ºC
ºC
|
|
Operating
Temp Range
Standard
Optional |
-20 to 70
-30 to 80
|
ºC
ºC
|
up to 95% non-condensing humidity
|
| Storage Temp |
-30 to 80 |
ºC |
|
| Size |
9.50 x 3.50 x 3.54
(24.1 x 8.9 x 8.98) |
inches
(cm) |
L x W x H |
| Weight |
6.5
(2.95) |
lb
(kg) |
|
| Splashproof Rating |
NEMA 13
|
|
|
| Acoustic Noise Rating |
160dB
|
|
|
| Shock Rating |
30
|
g
|
|
- Standard ranges listed, consult factory for special ranges.
- Maximum applied pressure without reducing sensor performance.
- For anticipated Line (Reference) Pressures . . . above 100 psi (689 kPa) on FS Pressure Ranges of less than or equal to 5 psi (34 kPa) or
- above 600 psi (4137 kPa) on FS Pressure Ranges of greater than 5 psi (34 kPa)
- calibration manifold should not be shifted at line pressure
- rate of line pressure change should be no more than 10 psi/sec
- Purge pressure must be greater than maximum input pressure.
- Static accuracy includes the combined errors due to non-linearity, hysteresis and non-repeatability.
- Includes thermal effects of zero & span relative to 25ºC.
- Consult factory for special temperature range calibrations.
Specifications subject to change without notice.
9116 Version 5.10 Release
Last Released Version: Version 5.09
Version 5.10 Changes
- Updated drivers to accommodate alternate memory parts.