You are here: Products  » Environmental » KPSI Selection Guide » KPSI 340

  Search
KPSI Series 340 ASIC-Based Small Bore Submersible Level Transducer   

The Series 340 is a small bore submersible hydrostatic level transducer that combines Pressure Systems’ excellent sensor competencies with the latest in Application Specific Integrated Circuit (ASIC) technology. Pressure Systems’ implementation of the ASIC provides unmatched sensor compensation over the entire operating range of the pressure sensor.

Series 340 Features

  • Custom Level Ranges up to 230 ft (70 m) H2O
  • Unparalleled Performance
       - Total Error Band: ±0.25% FS          
  • Analog Outputs of 4-20 mA or 0-5 VDC
  • Welded 316 SS or Titanium Construction
  • Small Diameter, Rugged, and Waterproof
  • Custom Cable Lengths

Series 340 Applications

  • Ground Water Monitoring
  • Surface Water Monitoring
  • Down Hole
  • Tailrace and Forebay Monitoring
  • Dewatering
  • Level Control

Buy Now

Support

Updates

Application Notes

FAQs

Warranty & Repair Policy

PDF Downloads

Datasheet

Manual

Accessories

Certifications

 

The Total Error Band specification (±0.25% FS) over the complete operating temperature range (-20 to 60ºC) eliminates the user having to combine multiple performance specifications to understand the total accuracy of the transducer.

All KPSI Transducers utilize a highly accurate pressure sensor assembly specifically designed for hostile fluids and gases. The assembly is integrated with supporting electronics in a durable waterproof housing constructed of 316 stainless steel or titanium. The attached electrical cable is custom manufactured to Pressure Systems’ specifications and includes Kevlar® members to prevent errors due to cable elongation, and a unique water block feature that self-seals in the event of accidental cuts to the cable. Each transducer is shipped with our latest SuperDry™ Vent Filter that prevents moisture from entering the vent tube for at least one year without maintenance, even in the most humid environments.

The Series 340 is CE compliant to EN 61000-6-4:2001 and EN 61000-6-2:2001 and have a IP 68 and NEMA 6P housing protection rating.  All KPSI Transducer calibrations are traceable to the National Institute of Standards and Technology (NIST).

Dated 04/09

Parameter
342
Units
Comments

LEVEL RANGES

Full Scale Level Ranges (1)  10 thru 230
(3 thru 70)

ft H2O
(m H2O)

for vented gage reference
35 thru 230
(10 thru 70)
ft H2O
(m H2O)
for sealed gage or absolute reference
Proof Pressure
1.5
x FS
 
Burst Pressure
2.0
x FS
 

STATIC PERFORMANCE

Total Error Band (2, 3)
±0.25

%FS TEB

Total Error from BFSL
Resolution
Infinitesimal

%FS

 

ENVIRONMENTAL

Wetted Materials 
316 SS or Titanium;
polyurethane or Viton®
   Viton® is a registered trademark of DuPont.
Compensated Temp Range 

-20 to 60

ºC
 
Operating Temp Range

-20 to 60

0 to 50

ºC

ºC

when using polyurethane cable

when using ETFE cable (4)

Protection Rating

IP 68, NEMA 6P

   

ELECTRICAL

Excitation

9-30

VDC

for mA and VDC output
Input Current

20
3.5

mA max

for mA output
for VDC output
Output

4 - 20
0 - 5

mA
VDC

 
Zero Offset

±0.12
< 0.1

mA
VDC

for mA output
for VDC output
Output Impedence

See Loop Resistance diagram page 5
510

ohm

for mA output
for VDC output
Insulation Resistance

100

mega ohm

at 50 VDC
Circuit Protection

Polarity, surge/shorted output

   

PHYSICAL

Approximate Weight
0.50 (224)
0.05 (79)

lbs (g)
lbs/ft (g/m)

transducer
cable
Cable
 
Jacket Material


  Pull Strength

  Number of Conductors
 
Conductor Size

Polyurethane (std)
ETFE (4) (opt)

200 (90)

4

22




lbs (kg)



AWG
Tefzel®, Teflon® and Kevlar® are registered trademarks of DuPont.
Cable Seal Molded polyurethane
Viton® Gland
  for polyurethane cable
for ETFE cable

LIGHTNING PROTECTION (OPTIONAL) (6)

Life Expectancy

>1000 operations

   
Peak Clamping Voltage 36 volts    
Response Time < 10 nsecs    
  1. Intermediate level ranges are available.
  2. Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability, and thermal effects over the compensated temperature range per ISA S51.1.
  3. TEB is based on averaging multiple readings.
  4. -20ºC to 50ºC for level ranges <= 100 ft (70m) H2O when using ETFE cable.
  5. ETFE is a fluoropolymer (Teflon® derivative) material, Tefzel® or equivalent.
  6. The power supply needs to be limited to 150 mA to avoid lock up of the gas tube after a suppression event.

Specifications subject to change without notice.


   © 2010 Pressure Systems Inc. Terms Of Use | Privacy Statement