Pressure Systems’ DTC Initium provides a powerful yet economical data acquisition system for PSI’s DTC series of miniature ESP pressure scanners. This advanced system epitomizes PSI’s ambition to eliminate in-situ calibration from electronic pressure scanning technology. The DTC Initium system integrates an advanced analog circuit design with PSI’s innovative Digital Temperature Compensation (DTC) technology to maintain optimal accuracy without on-line span calibration.
|

DTC Initium Features
- Support for 8 DTC Series ESP Scanners
- For up to 512 Pressure Channels
- ±0.05% FS Accuracy after Rezero
- 1200 Hz per Channel Throughput Capability in Engineering Units
- Auto-negotiating 10/100 Base-T Ethernet Interface
- Compact Robust Package
- Comprehensive Applications Utility Software
- Field Upgradeable Firmware via Website
DTC Initium Applications
- Windtunnel Aerodynamic Research
|
|
|
The proprietary technology utilizes a unique sensor mounting technique to enable accurate measurement of bridge resistance for use in real-time correction of thermal errors using personalized compensation algorithms embedded within the scanners. The result is an electronic pressure scanning system that provides state-of-the-art accuracy over the entire scanner operating temperature after only an initial rezero.
The Ethernet-based system directly connects any combination of up to eight 32 or 64 port DTC scanners to facilitate as many as 512 channels of pressure measurements. Multiple systems can be networked for larger channel requirements. The DTC Initium supports software and hardware triggering of data sets, data packet ID stamping for time correlation, and data throughput to the host computer at rates as high as 1200 Hz per channel in engineering units. The system also supports other sophisticated DTC scanner features including programmable deranging capability and internal manifold position sensing. The system operates from a single 18-36 VDC unregulated supply.
Windows® based utility software written in Visual Basic is supplied with the DTC Initium system allowing the user to configure acquisition parameters such as sample rates, averaging, and thermal update rates. It also provides diagnostic tools such as accuracy, channel crosstalk, and leak check routines. The utility software also facilitates firmware upgrades downloaded to the system via Ethernet, which are made available at no charge via PSI’s website.
Dated 09/08
| Parameter |
Specification
|
Units |
Comments |
|
CONFIGURATION
|
| Type of ESP's Supported |
ESP-64HD
ESP-32HD
|
|
DTC Series only |
| Number of ESP's Supported |
1 to 8
|
|
any combination of 32 or 64 port scanners |
|
PERFORMANCE (1)
|
System Static Accuracy (2)
@ 100% Range (3)
@ 33% Range (4) |
±0.05
±0.10
±0.25
±0.10
|
%FS
%FS
%FS
%FS
|
range >= 5 psid
range>= 10 in WC to < 5 psid
range < 10 in WC
range >= 5 psid |
Total Thermal Error
@ 100% Range
@ 33% Range |
±0.002
±0.004
±0.01
±0.003
|
%FS/ºC
%FS/C
%FS/ºC
%FS/ºC
|
range >= 5 psid
range >= 10 in WC to < 5 psid
range < 10 in WC
range >= 5 psid |
| A/D Resolution |
18 |
bits
|
|
| Measurement Resolution |
±0.003 |
%FS
|
|
Throughput Rate
(in Engineering Units) |
1200
650
325 |
Hz/ch
Hz/ch
Hz/ch
|
when scanning 16 channels per scanner
when scanning 32 channels per scanner
when scanning 64 channels per scanner |
|
COMMUNICATIONS
|
| Interface |
10/100 Base-T Ethernet |
|
auto-negotiating, half or full duplex |
| Protocol |
TCP and UDP |
|
|
|
ELECTRICAL
|
| Supply Voltage |
18-36 |
VDC |
unregulated |
| Supply Current |
1.6 - 0.8 |
A |
typically 1.2A @ 24 VDC with full load |
|
ENVIRONMENTAL
|
| Operating Temperature |
0 to 70 |
ºC |
|
| Storage Temperature |
-20 to +70 |
ºC
|
|
| Relative Humidity |
95 |
% |
at 50ºC (non-condensing) |
|
PHYSICAL
|
| Size |
10.00 x 7.00 x 3.50
(254 x 178 x 89)
|
in
(mm)
|
W x D x H |
| Weight |
3.6
(1.6) |
lb.
(kg.)
|
|
- User-selectable system operational parameters such as data averaging and sampling rates enable optimal measurement accuracy to be achieved.
- Static accuracy includes combined errors due to non-linearity, hysteresis, and non-repeatability.
- Achievable static accuracy after rezero. Scanner span calibration recommended at ambient temperature every 6 months to maintain
static accuracy specification.
- Accuracy is expressed as a percent of the deranged Full Scale when dual range calibration is specified.
Specifications subject to change without notice.