Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

Quartzonix™ Series 971 Intelligent Pressure Standard

Model 971 Intelligent Quartzonix™ Pressure Standards incorporate proven high accuracy vibrating quartz beam technology with low power miniature digital electronics to achieve a fully compensated 0.01% FS pressure measurement.

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 

Series 971 Intelligent Pressure Standard

Buy Now

 

Series 971 Features

  • Available Pressure Ranges:
    Barometric and
    0-15 through 0-500 psia
    (0-103 kPa through to 0-3447 kPa)
  • ±0.0001% FS Resolution
  • ±0.01% FS Accuracy
  • ±0.01% FS per 6 month Stability
  • Programmable Update Rates from 1 to 50 Hz
  • Power Consumption of only 250 mW
  • RS-485/RS-232 interface; programmable from 1200 to 115,200 baud

Series 971 Applications

  • Meteorological and barometric reference
  • Pressure calibration equipment

 

 


Specifications

–––––––––––––-------–
Updates
–––––––––––––-------–
Application Notes
–––––––––––––-------–

PDF Downloads
–––––––––––––-------–
Datasheet
–––––––––––––-------–
Manual
–––––––––––––-------– Certifications
–––––––––––––-------–

 


 
 

Model 971 Intelligent Quartzonix™ Pressure Standards incorporate proven high accuracy vibrating quartz beam technology with low power miniature digital electronics to achieve a fully compensated 0.01% FS pressure measurement.  Quartzonix™ pressure standards use a patented quartz structure and vibrating beam to measure pressure-induced stress.  The beam’s frequency of vibration changes by pressure induced stress that is applied to the beam via a miniature metal bellows.  The bellows isolates the quartz sensing element from the applied gas, thereby making the sensor insensitive to gas density.  The Model 971 Intelligent Quartzonix™ has measurement resolution of up to 0.00001% FS, and drift rates of less than 0.01% FS over a 6-month period.

The Model 971 uses a full duplex RS-232 or RS-485 type interface with programmable baud rates from 1200 to 115,200. Utilizing a multi-drop protocol, a network of up to 31 transducers can be configured on the same bus. The measurement units and the update rate (from 1 to 50 readings per second) are both user-programmable. The extremely low power consumption design with a programmable sleep mode makes the Intelligent Quartzonix™ ideal for remote battery power applications. Firmware upgrades providing new features and capabilities are available on Pressure Systems’ website at no charge and can be downloaded into the 971 via the serial interface.


Dated 07/07

 
 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Register | Search