Parameter
|
9400
|
9401
|
9402
|
Units
|
Comments
|
PNEUMATICS |
| Pressure Reference |
Gauge (1)
|
Absolute
|
Differential
|

|
 |
| Pressure Ranges |
9400 Gauge
5 (34)
10 (69)
15 (103)
50 (345)
100 (689)
250 (1724)
500 (3447)
750 (5171)
1500 (10342)
3000 (20684)
5000 (34474)
10000(68947)
|
9401 Absolute
15 (103)
30 (207)
50 (345)
100 (689)
250 (1724)
500 (3447)
750 (5171)
1500 (10342)
3000 (20684)
5000 (34474)
10000(68947)
|
9402 Differential
5 (34)
10 (69)
15 (103)
50 (345)
100 (689)
250 (1724)
|
psi (kPa) |
Contact factory for other ranges.
|
| Proof Pressure (2)
|
3.0 1.5 |
x FS
|
range less than or equal to 100 psi (689 kPa)
range
greater than 100 psi (689 kPa) |
| Burst Pressure
|
5x 3x 2x |
x FS
|
range equal to 5 - 500 psi
range
greater than 500 - 3000 psi
range
greater than 3000 psi |
STATIC PERFORMANCE |
Static Accuracy (3) |
±0.05
±0.25
±0.50 |
±0.5
N/A
N/A |
% FS % FS
% FS |
range less than 750
psi
range greater than 750 psi to less than equal to 3000 psi
range greater than 3000 psi |
Total Thermal Error (4) |
±0.005 ±0.05
|
%FS/șC %FS/șC
|
digital compensation
(6)
analog compensation (7) |
| Thermal Hysteresis |
±0.2 |
% FS |
after cycling over
full temp range |
| Max Line Pressure |
N/A |
1000 (6885) |
psi
(kPa) |
|
| Line Pressure Effect (5) |
N/A |
±0.01 |
%FS/psi |
|
ENVIRONMENTAL |
| Wetted Materials |
316 SS & Viton
|
 |
 |
| Compensated Temp Range
|
0 to 50
-30 to 40
|
șC
|
standard
optional |
| Operating Temp Range |
-30 to 100
|
șC
|
 |
ELECTRICAL |
| Excitation |
12 - 36
|
VDC |
 |
| Power Supply Rejection |
±.001
|
%/VDC |
 |
Output Vo Vt |
0 - 4.9
3.0 - 4.0
|
VDC VDC
|
 |
Output Impedance Vo Vt |
1000 1000
|
Ohm Ohm
|
max max |
| Insulation Resistance |
50
|
M Ohm |
@ 50 VDC |
PHYSICAL/ENVIRONMENTAL |
| Acceleration |
±0.02
±0.01 |
% FS/g
% FS/g |
Range less than equal to 15 psi (103 kPa)
Range greater or equal to 30 psi (207 kPa) |
| Vibration |
±0.05 |
% FS/g |
30 g peak 10 Hz - 2kHz |
| Weight |
6 (170) |
17 (480) |
oz. (gm) |
 |
| Pressure Connection |
1/4" NPT, Male
AN4, Male
G 1/4, Male
M14, Male
|
7/16"
Female SAE
o-ring Boss |
 |
|
Electrical Connection Standard Optional |
PTIH-12-8P Bendix
PVC Jacketed Cable |
 |
 |
|
(1) |
Pressure ranges greater than 750 psi are "sealed"
gage rather than "vented" gage. |
| (2) |
Maximum pressure which can be applied without causing calibration shift. |
| (3) |
Static accuracy includes
the combined errors due to nonlinearity, hysteresis and nonrepeatability
following a zero calibration of the sensor. |
| (4) |
Includes effects of zero and span relative to 25șC.
|
| (5) |
Primarily zero offset.
|
| (6) |
For digital compensation
type transducers, the on-board EEPROM contains full operating range temperature
and nonlinearity correction coefficients. |
| (7) |
For analog compensation
type transducers, the on-board EEPROM contains room temperature (approx. 25șC)
nonlinearity correction coefficients. |
|
|
|
Specifications subject to change without notice
|