Parameter
9021
Units
Comments

PRESSURE CHARACTERISTICS

Number of
Transducers Supported
12

channels

pressure transducers or voltage inputs
STATIC ACCURACY AND RESOLUTION
Measurement Resolution
±0.003
% FS

Anti-Aliasing Filter 30 Hz cutoff frequency
4pole Bessel filter on pressure
and temperature signals
Voltage Measurement Accuracy (1)
±0.025
±0.025
±0.05
±0.05
% FS
% FS
% FS
% FS
0 to 4500 mV
0 to 250 mV
0 to 100 mV
0 to 50 mV
Thermal Error ±0.001 % FS/ºC when operating outside the calibrated temp range using FS voltage ranges </= 250 mV

DATA SCAN & TRANSFER RATES

Measurement Scan Rate
100
meas/ch/sec

COMMUNICATION

Ethernet 10Base T   half duplex
Protocol TCP and UDP   static or dynamic IP assignment

POWER/SIGNAL REQUIREMENTS

Input Voltage
18 to 36 
VDC unregulated supply
Input Current
330
mA max ² @ 24VDC
Hardware Trigger
Threshold
2.5
VDC TTL compatible diff input, ±5
VDC common mode

TRANSDUCER EXCITATION VOLTAGE CHARACTERISTICS

EXCITATION Voltage
5 or 10  
VDC user selectable, 10 VDC default
Excitation Accuracy ±0.07 % @ 25ºC
Temperature Coefficient 20 ppm/ºC  
Long term Stability 15 ppm measured over 24 hours
Load Regulation 0.3 % 100mA max total load

ENVIRONMENTAL/PHYSICAL

Calibrated Temp Range 
0 to 50
șC
consult factory for extended temps
Operating Temp Range
-30 to 70
șC
< 95% non-condensing humidity
Storage Temp
-30 to 80
șC
Size
9.5 x 3.5 x 3.4
(24 x 9 x 8.7)
inches
(cm)
L x W x H
Weight 4.3
(1.95)
lb
(kg)
 
(1) For channels measuring voltage inputs, measurements made from -FS to 0 VDC are specified at twice the error specification for the related range.
(2) Worst case at maximum tarnsducer load.
 

 

Specifications subject to change without notice.

 

Refer also to Series 9400 Digitally Compensated
All-Media Pressure Transducer datasheet for transducer specifications.

 
 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search