Esterline Pressure Systems Logo header spacer Pressure Systems Site Map Pressure Systems Home
 
 

NetScanner™ Model 9016 Ethernet Intelligent Pressure Scanner

The Model 9016 Pneumatic Intelligent Pressure Scanner is a completely self-contained pressure acquisition module for multiple measurements of dry, non-corrosive gases. The scanner integrates 16 silicon piezoresistive pressure sensors of any pressure range with a microprocessor in a compact, rugged package.

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
 

9016 Intelligent Pressure Scanner

 

Model 9016 Features

  • ±0.05% FS System Accuracy
  • EU Throughput Rates of 100 Hz
  • 10 BaseT Ethernet with TCP & UDP Protocol
  • Pressure Ranges up to 850 psid (5860 kPa)
  • Built-in Rezero, Purge and Leak Check Features
  • Triple Scan List Capability
  • Rugged Splashproof Enclosure
  • Hardware and Software Triggered Data Acquisition

Model 9016 Applications

  • Turbomachinery Test Stands
  • Vehicle Test

 

 


Specifications

–––––––––––––-------–
Updates
–––––––––––––-------–
Application Notes
–––––––––––––-------–
Software / Firmware
–––––––––––––-------–
FAQ

PDF Downloads
–––––––––––––-------–
Datasheet
–––––––––––––-------–
Manual
–––––––––––––-------–
Quick Start
–––––––––––––-------–
Upgrade Instructions
–––––––––––––-------–
UDP Query

 


 
 

Each field-replaceable pressure sensor incorporates a temperature sensor and EEPROM for storage of calibration data as well as sensor identification information such as pressure range, factory calibration date, and user-managed last or next calibration date.

The microprocessor uses the data from the EEPROM to correct for sensor zero, span, linearity and thermal errors. It also performs digital temperature compensation of the sensors to reduce thermal errors by a factor of ten or more over conventional sensor compensation. The microprocessor also controls the actuation of an integral manifold to execute on-line zero or span calibration upon request. This capability virtually eliminates sensor zero drift error and provides guaranteed system accuracy of up to ±0.05% FS (Full Scale) after rezero.

The 9016 can sample using up to three concurrent scan lists at continuous rates up to 100 measurements per channel per second in engineering units. Pressure data in engineering units is output through a 10 Mbit Ethernet interface supporting both TCP and UDP protocols. The scanner is supplied with comprehensive Visual Basic-based software called NUSS for PC compatible computers. Firmware upgrades are made available at no charge via PSI’s website and can be downloaded to the scanner over the Ethernet interface using NUSS.

The Model 9016 Intelligent Pressure Scanner is one component of the NetScanner™ System. Multiple NetScanner components measuring a variety of parameters and sharing the same command set can be networked to form a distributed intelligent data acquisition system.

Dated 09/04

 

 
 
Pressure Systems
34 Research Drive
Hampton, VA 23666
800.678.7226
KPSI Selection Guide Quality Policy
ISO-9001:2000
Receive E-Newsletter
IPC Member Logo WEEE/RoHS Statement
WEEE/RoHS
Statement
 


Home | Environmental Products | Industrial Products | Aerospace Products | Buy Now
KPSI Level and Pressure Selection Guide | Quartzonix Pressure Standard | NetScanner System | System 8400 |
Electronic Pressure Scanners | S8256 System | Miscellaneous Hardware | News Archive
Contact Us | Distributors | Links | Employment | Register | Search