Parameter |
552
|
553
|
Units |
Comments |
LEVEL RANGES |
| Full Scale
Level Ranges (1) |
10 thru 692
(3 thru 210) |
ft H2O
(m H2O) |
vented gage reference |
10, 35, 50, 70, 230, 692
(3, 10, 15, 21, 70, 210)
|
ft H2O
(m H2O) |
absolute reference |
| Proof Pressure |
1.5 |
x FS |
 |
| Burst Pressure |
2.0 |
x FS |
 |
MEASUREMENT ACCURACY |
| Level (2, 3) |
±0.05
±0.10
|
%FS TEB
%FS TEB
|
for level ranges > 10 ft (3 m) H2O
for level ranges = 10 ft (3 m) H2O |
| Temperature |
±1.0
|
±0.2
|
ºC |
|
DATALOGGING SPECIFICATIONS |
| Modes of Sampling |
Linear, Linear Avg, Event, Logarithmic |
|
software selectable |
| Pressure Units Available |
psi, ft H2O, mm H2O, cm H2O,
m H2O, kPa
|
|
user-specified also available by entering slope
and offset |
| Sampling Rate (4) |
Programmable, 15 readings per second max |
|
|
| Internal Non Volatile Memory |
2 |
4 |
Mbyte |
|
| Maximum Level Scans |
196,560 |
393,120 |
|
Level with Time-Stamp |
| Maximum Level and Temperature Scans |
144,144 |
288,288 |
|
Level and Temperature with Time-Stamp |
| Time-Stamp Accuracy |
±2 |
min/year |
over the compensated temp range |
| Data
Upload Time |
100
|
scans/sec |
(Level and Temperature with Time-Stamp) per
second @ 19200 baud |
ENVIRONMENTAL |
| Wetted Materials |
316 SS or Titanium; Viton® |
 |
Viton®is a registered
trademark of DuPont. |
| Compensated Temp Range |
-10 to
40
|
º
C
|
 |
| Operating Temp Range |
-20 to 50
|
º
C
|
 |
| Protection
Rating |
IP 68, NEMA 6P |
|
|
| Storage Temp Range |
-40 to 80 |
ºC |
|
ELECTRICAL |
| Excitation (5) |
6-16 |
VDC |
|
| Current Draw |
8.0
25.0
0.25
|
mA
mA
mA
|
average current during
measurement
15 mS
peak current during page writes
quiescent |
| Communication Interface |
RS-485
|
|
3 volt
p-p differential with selectable baud rates from 1200 to 19200 |
| Protocol |
SDI-12 |
|
ver 1.3 compliant |
PHYSICAL |
| Approximate Weight |
0.60 (272)
0.05 (79) |
lbs (g)
lbs/ft (g/m) |
transducer
cable |
Cable
Jacket Material
Pull Strength
Number of Conductors
Conductor Size |
Polyurethane (std)
Tefzel® (opt)
200 (90)
4
22 |
lbs (kg)
AWG |
Tefzel®, Teflon® and
Kevlar® are registered trademarks of DuPont.
|
| Cable Seal |
Molded Polyurethane
Viton® Gland |
|
for polyurethane
cable
for Tefzel® cable |
| (1) |
Intermediate level
ranges are available. |
|
(2) | Total Error Band (TEB) includes the combined errors due to nonlinearity, hysteresis, nonrepeatability and thermal effects over the compensated temperature range per ISA S51.1. |
| (3) |
For absolute reference, additional error up to ±0.2% FS/year may be incurred due to offset drift. |
| (4) |
For sampling rates greater than 3 Hz, the pressure/temperature averaging must be reduced as required. |
| (5) |
An external battery with 2300 mAh capacity, sampling 4 times/hour, will allow datalogging for 1 year. |